The TURBILIGHT II® is designed for fast and reliable automated measurement of medium and low turbidity (0-2 to 0-1000 NTU) mainly in drinking and surface water. Easy to install, operate and service, endowed with an efficient self-cleaning dispositive, The TURBILIGHT II® is designed for online monitoring of turbidity suited to a large range of applications.
Analytical Performances
Technology Continuous measurement by IR Nephelometry
Accuracy +/- 2 % full range
Repeatability +/- 2 % full range
Resolution 0.001 NTU for 0-2 NTU MR
Measuring range Programmable 0-50 / 100 / 250 / 500 / 1000 NTU (0-4000 mg/l on request)
Response time Continuous online monitoring Initial response time (cold start) in few seconds – T90<20sec.
[/av_iconlist_item]
[av_iconlist_item title='Sampling System' link='' linktarget='' linkelement='' icon='ue8da' font='entypo-fontello' av_uid='av-5e01p6k']
Cell cleaning Measuring cell equipped with an automatic cleaning system – Piston operated wiper. User programmable frequency.
Sequential control 1 to 5 tanks / purge valves with 1 single electronic cabinet / 1 single capacitive probe
Sample inlet
Temperature +4 to + 40C
Flow 50 l/h
Pressure 0.1 to 3 bar
Connections
Sample inlet: semi-rigid tubing 6x8mm
Sample outlet: flexible tubing 12 x 14mm